pH sensor

Choose one of many pH Single-Use-Sensor's pre-installed

Various pH Single-Use-Sensors (SUS) is custom designed to be integrated along with SUB or SUF assembling. Before bagging and subsequent E-beam irradiation. Dry storage capability are 12 month on all pH SUS stored below 20°C and with no UV light exposure. 

OneFerm™ with S8 or K8 socket, classical analogue signal, 1-13 pH range, 12 mm body, zero drift. Standard format is PG 13,5 x 120, 225, 325 mm, 425.

 

OneFerm™ VP socket, classical analogue signal, built in Pt100 temperature sensors, connector VP6 / VP8, 1-13 pH range, 12 mm body, zero drift. Standard format is PG 13.5 x 120, 225, 325, 425 mm.

 

OneFerm™ VP socket, classical analogue signal, built-in NTC temperature sensor, designed for connection with the ARC transmitter for digital communication. Standard format is PG 13.5 x 120, 225, 325, 425 mm.

 

Glass pH analogue SUS need 10 minute conditioning in media before one point calibration and use. Glass tube will turn brown after E-beam exposure. Handles steam sterilization for multiple use though we do not guaranty the performance.

 

Specification Sheet is available to download from Hamilton

FermProbe™ with S8 socket, classical analogue signal, 1-13 pH range, 12 mm body, zero drift. Standard format is PG 13.5 x 120, 225, 325 mm.

 

  • Classical analogue signal and S8 socket.
  • Polymer pH analogue SUS need 10 minute conditioning in media before one point calibration and use.
  • Perform identical to the traditional glass Re-Usable-Sensors. 
  • Not designed for steam sterilization.  
  • An adapter from S8 to (cylindrical) K9 socket is available. Check out adaptors.

Specification Sheet and manuals available to download from Broadley James Corp

Presens pH spot sensor now available pre-installed in different length as to the chosen SUB, SUF, SUM size and brand. Even the BioBLU can be customized and include the Presens pH sensor fitting the BioFLO PCS.